| Accuracy (Grade) | Maximum Permissible Error | |
| 0.05 | ±0.05% lower limit of the measurement range (when the pressure is below the lower limit of the measurement range) | ±0.05% actual measurement range (when the pressure is within the measurement range) | 
| Model | Unit | TV-6T | |
| Measurement Range | 0.04~0.6 | ||
| Upper limit | MPa | 0.6 | |
| Upper limit | MPa | 0.04 | |
| Upper limit | MPa | 0.6 | |
| Upper limit | MPa | 0.04 | |
| Nominal area of piston | cm2 | 1 | |
| Bedplate & piston | Nominal mass | Kg | 0.4 | 
| Pressure | MPa | 0.04 | |
| Specific weight | Nominal mass | Kg | 0.1; 0.5 | 
| Pressure | MPa | 0.01; 0.05 | |
| Quantity | unit | 6; 10 | |
| Joint | M20×1.5 | M20×1.5 | |
| Weight | Kg | 35 | |
| Work medium | Voltage transformer oil or the mixed oil (voltage transformer oil and kerosene oil) when 20℃, the CSt is 9~12mm2/s; Acid value:≤0.05KOHmg/g. | ||